Images are representations only.

SITRANS TH200 Temperature Transmitter

Brand: Siemens AG
Does not ship to Virginia, United States
SITRANS TH200 is designed to support all common RTDs, thermocouples, resistance and millivolt sensors.

Setup is quick and easy with the transmitter modem and SIPROM T software.

Unique user-friendly details are implemented, one of them being the direct readout of the output current with a multimeter without having to open the 4 to 20 mA loop.

A red/green LED indicates the device status at a glance.

Transmitters of the "intrinsically safe" type of protection can be installed within potentially explosive atmospheres.

The devices meet the directive 2014/34/EU (ATEX) as well as the FM and CSA requirements.
Features
  • Measures temperature with a linear output signal
  • High accuracy across entire ambient temperature range
  • Reduces plant noise
  • Alarm signal for sensor break or short-circuit according to NAMUR 43
  • Electro magnetic compatibility (EMC) according to DIN EN 61326 and NE21
  • Saves expensive compensation cables with its internal reference junction
  • Galvanic insulation for accuracy and safety in thermocouple applications
  • Single- or dual-point trim
  • Custom characteristic curve for application of non-standard sensors
  • Difference- and average- measurement
  • Explosion protected variants for US, Canada and Europe
  • Rugged design, fully potted electronics
show more
SITRANS TH200 is designed to support all common RTDs, thermocouples, resistance and millivolt sensors.

Setup is quick and easy with the transmitter modem and SIPROM T software.

Unique user-friendly details are implemented, one of them being the direct readout of the output current with a multimeter without having to open the 4 to 20 mA loop.

A red/green LED indicates the device status at a glance.

Transmitters of the "intrinsically safe" type of protection can be installed within potentially explosive atmospheres.

The devices meet the directive 2014/34/EU (ATEX) as well as the FM and CSA requirements.
Features
  • Measures temperature with a linear output signal
  • High accuracy across entire ambient temperature range
  • Reduces plant noise
  • Alarm signal for sensor break or short-circuit according to NAMUR 43
  • Electro magnetic compatibility (EMC) according to DIN EN 61326 and NE21
  • Saves expensive compensation cables with its internal reference junction
  • Galvanic insulation for accuracy and safety in thermocouple applications
  • Single- or dual-point trim
  • Custom characteristic curve for application of non-standard sensors
  • Difference- and average- measurement
  • Explosion protected variants for US, Canada and Europe
  • Rugged design, fully potted electronics

Specifications

Brand

Output Signal

  • 4-20 mA, 2-Wire

Auxiliary Power

  • 11-35 Vdc

Load

  • 500 ohms

Error Due to Internal Reference Junction

  • < 0.5° C

Ambient Temperature

  • -40° C to 85° C
  • -40° F to 1855° F

Electromagnetic Compatibility

  • acc. to EN 61326 and NE21

Relative Humidity

  • < 98 %, with condensation

Material

  • Molded Plastic

Weight

  • 50 g

Degrees of Protection

  • Enclosure: IP40
  • Terminals: IP00

Certifications & Approvals

  • PTB 05 ATEX 2040X
  • II 1 G Ex ia IIC T6/T4
  • II 2 (1) G Ex ia/ib IIC T6/T4
  • II 3(1) G Ex ia/ic IIC T6/T4
  • II 1D Ex iaD 20 T115° C
  • II 3 G Ex nL IIC T6/T4
  • II 3 G Ex nA IIC T6/T4
  • FM 3024169
  • FM 3024169C
  • IS / Class I, II, III / Div. 1 / GP ABCDEFG T6, T5, T4
  • EAC Ex(GOST)
  • NEPSI
  • IEC
  • EXPO-LABS

Units of Measure

  • °C
  • °F

Sensor Current

  • < 0.45 mA

Response Time

  • < 250 milliseconds for 1 sensor with break monitoring

Sensor Type

  • Pt100 (IEC) 2-Wire
  • Pt100 (IEC) 3-Wire
  • Pt100 (IEC) 4-Wire
  • Thermocouple Type: B, C, D, E, J, K, L, N, R, S, T, U

Typical Applications

  • SITRANS TH200 transmitters can be used in all industrial sectors.